Magstim BiStim²🔗

The BiStim² is a paired-pulse stimulator combining 2 monophasic stimulators to deliver paired pulses through 2 separate or through a single coil. The paired pulses can be delivered with independent stimulation intensities and variable inter-pulse intervals.

Magstim devices share common features that are detailed in the common features section of the guidelines.

../_images/bistim-system.png

Start-up🔗

The BiStim² can be started by (1) setting the power switch to ON behind both units, and (2) pressing the power button on the front panel of both units.

Note

Contrary to the Super Rapid² Plus¹, the BiStim² power consumption is handled through a single power outlet.

Coils🔗

TODO

To use the BiStim² as a paired-pulse stimulation with a single coil, the 2 stimulators must be connected to the appendix on the left side of the stimulator A, and the coil must be connected to the same appendix.

../_images/bistim-coil-connection.png

Important

If the coil(s) are connected to a stimulator directly, the stimulators work independently one from another.

Settings🔗

When both stimulators are connected to the same coil, the stimulator A controls the intensity of both pulses (and stimulators) and the inter-pulse interval. The pulses can be set to different intensities.

Press on the nob to circle between the 3 settings and turn the nob to set the desired value.

../_images/bistim-settings.png

Intensity is set as a percentage of the maximum stimulator output. The intensity should be defined based on the individual motor threshold.

The inter-pulse interval is set in milliseconds and defines the delay between the pulse delivered by both stimulators.

Attention

Using both BiStim² stimulators connected to the same coil with an inter-pulse interval set to 0 is not equivalent to using a single BiStim² stimulator. This mode of operation is not recommended.

When setting the inter-pulse interval to 0, then pressing the red button and turning the nob counter-clockwise, the BiStim² will enter mode ε, displaying an E in the inter-pulse interval display. In this mode, the pulse delivery is controlled by the triggers only. The first trigger will deliver a pulse on stimulator A, the second trigger will deliver a pulse on stimulator B.

../_images/bistim-settings-epsilon.png

Tip

To exit the mode ε, press the red button and turn the nob clockwise.

Important

The BiStim² stimulators need time to recharge between pulses. Thus, after a pulse has been delivered on a stimulator, it is up to the user to ensure no new trigger attempts to deliver a pulse on the same stimulator before it is ready. Usually, the stimulator requires around 1 second to recharge (depends on the pulse intensity).

If a trigger is sent to a stimulator that is not ready, the BiStim² will wait until it is recharged before delivering the pulse. Thus, the inter-pulse interval will not be respected.